Sirona Dental Systems GmbH
6 Checklist
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
6.2 Construction requirements
båÖäáëÜ
6.2
Construction requirements
Electrical connections for Teneo, Sinius
Uneven floors for Teneo, Sinius
Connections: Media (see On-site installation [ → 9])
☑
● Water supply
Pipe 10x1 mm, corner valve outlet 3/8"
☐
● Compressed air supply line
Pipe 10x1 mm, corner valve outlet 3/8"
☐
● Suction line
DN 40 HT-PP ISO 8283-3, inner diameter 36.5 mm
☐
● Water drainage
DN 40 HT-PP ISO 8283-3, inner diameter 36.5 mm
☐
● Installation pipe (power supplies)
min. DN 40 HT-PP ISO 8283-3, inner diameter 40 mm
☐
● Installation pipe (IT)
DIN 40 HT-PP ISO 8283-3 (or corresponding flat conduit)
☐
Connections: Electrical (see On-site installation [ → 9])
☑
● Power cable: 3 x 1.5 mm
²
☐
● Type B automatic circuit breaker
230 V AC, 16 A slow-blow or
● Type B automatic circuit breaker
100-115 V AC, 20 A slow-blow
☐
☐
● Suction machine control cable and call cable: 3 x 1.5 mm
²
☐
● Wireless systems in 2.4 GHz frequency range available?
(e.g. room monitoring systems, video transmitters, etc.)
☐ Yes
☐ No
● If yes, this may damage the wireless foot control.
Please consult our Product Service team.
Uneven floors: Mounting plates (see Mounting plates [ → 38])
☑
● No plate required (unevenness max. 2 mm, shimming plates can be used)
☐
● Adapter plate, REF 63 07 511
☐
● Demo plate (demo operation at a trade show), REF 63 08 717
☐
● Load capacity of the floor is indicated (pressure and tensile loads).
☐
● The treatment center can be safely anchored in the load-bearing structure (concrete/wood;
NOT screed).
☐