63 22 668 D3561
42
D3561
.
021.01.06
.
02 12.2014
4 Dimensions, technical data
Sirona Dental Systems GmbH
4.6 Technical data
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
Operating and transport conditions
Approval and year of manufacture
Transport and storage
conditions:
Temperature: -40 °C – +70 °C
(-40 °F – 158 °F)
Relative humidity: 10% – 95%
Air pressure: 500 hPa – 1,060 hPa
Operating conditions:
Ambient temperature: 10 °C – 40 °C (50
°F – 104 °F)
Relative humidity: 30% – 85%
no condensation
Air pressure: 700 hPa – 1,060 hPa
Installation location:
≤ 3000 m above sea level
This treatment center is not suitable for
operation in areas subject to explosion
hazards.
Pollution degree:
2 acc. to IEC 60664-1
-40
+70
1060
500
95
10
+10
+40
1060
700
85
30
Tests/Approvals:
See "Standards/Approvals“ [ → 44].
Year of manufacture:
(on the rating plate)
USB port:
corresponds to USB 2.0 standard
20XX