63 22 668 D3561
16
D3561
.
021.01.06
.
02 12.2014
3 On-site installation
Sirona Dental Systems GmbH
3.4 Underfloor installation of supply lines
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
1
Water supply
Pipe 10x1 mm, corner valve outlet 3/8"
2
Compressed air supply
Pipe 10x1 mm, corner valve outlet 3/8"
3
Suction line
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter 36.5 mm)
4
Water drainage
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter 36.5 mm)
5
Installation pipe
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter 40 mm)
6
Control cable to relays for the
suction machine (
),
call cable ( ),
Special function (#)
3 x 1.5 mm
2
(quality as in the power cable)
7
Power supply
3x1.5 mm
2
Circuit breaker: for 230 VAC: 16 A slow-blow
for 100 - 115 VAC: 20 A slow-blow
Recommended: Type B automatic circuit breaker
8
not connected
9
Installation pipe, inner diameter 50 mm
(or corresponding flat conduit) for the PC connection
%
Содержание Sinius
Страница 1: ...f o NOKOMNQ pfkfrp L pfkfrp p L pfkfrp qp k W b Installation requirements...
Страница 2: ......
Страница 57: ......