Instruction Manual FireGuard 2
Troubleshooting
86
13542E/4
The following warning messages can be displayed:
WARNING
MESSAGE
DESCRIPTION
POSSIBLE CAUSES
V IN
The input voltage is outside
the permitted range (9 to 30
VDC).
The service voltage is faulty
ADJUST FAULT
Recalibration could not be
carried out.
The instrument is soiled
The nominal value for the ad-
justment does not match the
value of the checking unit
SOILING
The soiling level is greater
than the set limit.
The measuring cell is soiled and
has to be cleaned
The cleaning interval is too
long
The installation position is in-
correct
FLOW RATE
The flow rate monitoring has
been actuated. In the last 4
days, the value of the turbidity
gradient was always smaller
than the flow rate limit.
No air flow in tunnel
No traffic
The foreign body protector is
extremely soiled
The optics are extremely soiled
HEATER
The sample heater does not
reach its nominal tempera-
ture.
Very cold environment with
high winds
The heater is defective
OVER TEMP
The temperature in the in-
strument has exceeded 65 °C.
The medium or ambient tem-
perature is too high and defec-
tive/no cooling
CURRENT 1 .. 2
Current output 1 .. 2 is dis-
turbed.
Terminals are open
Interruption of the current loop
of the reading output
TEMP.SENSOR
The inner temperature sensor
failed.
Defect in the electronic system
Service technician
SERVICE
Shows when service work is
due.
Service work is due.
VERS.SD CARD
The data on the microSD card
does not match the current
software.
Update was performed incor-
rectly.
WATCHDOG
The internal fault monitoring
has been actuated.
The program has been re-
started.
Program crash
Table 3: Possible warning messages