Instruction Manual FireGuard 2
Servicing
76
13542E/4
WORKSTEP
ADDITIONAL INFO / IMAGES
5.6: Insert the heating elements (B) into the in-
sulating shell (A).
5.7: Reinsert the sample heaters (Figure 31,
pos. 4) into the measuring cell inserts (Figure
31, pos. 3).
6.
Clean the inside of the measuring cell housing
(Figure 31, pos. 2).
Use a brush and damp cloth for cleaning
(no solvents!)
7.
Assemble the measuring cell housing:
Put the two measuring cell inserts together
(Figure 31, pos. 3) and insert into the measur-
ing cell housing (Figure 31, pos. 2).
The cam on the measuring cell housing (arrow)
must be aligned to the continuous slot on the
measuring cell insert.
Check that the two white points (circle) are
aligned to one another.