Instruction Manual FireGuard 2
Operation
50
13542E/4
Pos : 106 /Bedi enung/I n den Ser vic ebetrieb umschalt en @ 2\ mod_1383915568847_20.doc x @ 13702 @ 2 @ 1
7.11
Switching to service mode
The system is configured in service mode. The measuring procedure is interrupted and the
main menus appear on the display. Service mode is accessed as follows:
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
The main menus appear.
The instrument is now in service
mode.
The following applies in service mode:
* The measuring values remain on the last values on the digital interfaces.
* Depending on the configuration, the current outputs go to 0/4 mA or remain on the
last measuring values.
The limits are deactivated.
If an output for service is programmed, it is activated.
Fault messages are suppressed.
* This does not apply when the
Current outputs\General\For service
parameter is set to
Measure
.
For measuring mode, press the
Meas
button. When switching from service mode to meas-
uring mode, an hourglass appears in the information bar for about 10 seconds. The meas-
uring values are frozen during this time.