Ionics Instruments
2004
Page 3-12
DLM 30007-08 Rev. A
The TOC analyzer can also be disconnected from the stream selector for
continuous monitoring of a single sample stream, measurement of grab
samples or calibration of the analyzer.
Three multi-stream systems are available: a 120V system for the Model 800,
a 100V system for the Model 810 and a 230V system for the Model 820.
Turbo
Turbo Sampling Mode is an option on the analyzer achieved through specific
firmware. Analyzers configured with this option are designated as the “Turbo”
version of the model number (e.g., 800 Turbo analyzer). This mode is useful
for monitoring recycle or reclaim water in semiconductor facilities where rapid
response is of primary concern. When operated in the Turbo mode, the
analyzer responds to a step change with an initial response in 3 minutes and
90% of final value in 3.5 minutes. Results are reported every 3 to 4 seconds.
Flow Switch and Binary Input Module
For on-line applications that experience intermittent water flow, the Flow
Switch along with the Binary Input Module allow the analyzer to be turned on
and off when water flow starts and stops. Ionics Instruments has tested the
flow switch for functionality and ease-of-installation. Flow Switches must be
used with a prefilter, and water must flow from below the analyzer before
proceeding horizontally to the analyzer. The Flow Switch will not operate with
the Multistream.
Ozone Destruct Kit
This kit is designed to easily replace the analyzer’s existing sample line with ozone
destruct tubing. Once installed the tubing will eliminate up to 800ppm of ozone from
ozonated water systems for five years of operation.
Environmental Enclosure
Two Sievers enclosures are offered: the Protective Enclosure and the
Environmental Enclosure. These enclosures protect the analyzer in corrosive,
or extreme temperature environments:
Содержание TOC 800
Страница 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Страница 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Страница 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Страница 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Страница 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...