Ionics Instruments
2004
Page 8-16
DLM 30007-08 Rev. A
WARNING
The procedures described below reset timers and other key indicators used to
keep track of the Reagents, UV Lamp, and Pump Tubing lifetimes. In these
options, pressing ENTER zeroes the timers. Do not reset the times unless the
consumables are replaced. When viewing the status of the Reagents, UV Lamp
and Pump Tubing, press CLEAR to exit the menu without resetting the timers.
Oxidizer Reagent
To enter the installation date for the oxidizer reagent supply:
1. Use the arrows ( or ) to scroll to OXIDIZER. Press ENTER to
display the OXIDIZER INSTALLATION menu:
ENTER
CLEAR
INSTALL
SET % LEFT
SET INSTALL DATE
FIGURE 8-6: OXIDIZER INSTALLATION Menu
2. With the INSTALL option highlighted, press ENTER to display the
OXIDIZER Status and INSTALLATION screen: