Ionics Instruments
2004
Page 6-16
DLM 30007-08 Rev. A
valve, available through Ionics Instruments to eliminate the need to change the
plumbing when switching between on-line and grab sampling.
Setting the Sample Mode to Grab Sampling
To set the sample mode to Grab Sampling from the MAIN menu:
1.
Use the arrows ( or ) to scroll to SETUP and press ENTER.
The
SETUP menu is displayed, with the CLOCK function highlighted.
CLOCK
REAGENTS
PRINTER
OUTPUTS
ENTER
CLEAR
UV LAMP
PRESETS
SAMPLE MODE
PASSWORD
FIGURE 6-19: SETUP Menu
2. Use the arrows ( or ) to scroll to SAMPLE MODE. Press ENTER to
display the SAMPLE MODE menu.
ENTER
CLEAR
ENT = SELECT, CLR = EXIT
ONLINE SAMPLING
CURRENT MODE: ONLINE
GRAB SAMPLING
FIGURE 6-20: SAMPLE MODE Menu