Ionics Instruments
2004
Page 10-6
DLM 30007-08 Rev. A
pump labeled “Acid” and “Oxidizer”, and make sure the connectors are firmly
in place. Otherwise, contact Ionics Instruments.
Warning 35:
The clock has been set to a time earlier than the time of data
written on the RAM card. For example, this error will occur when the clock is
changed from daylight savings time to standard time. This error will be
repeated until the clock catches up to the latest time/date written on the RAM
card. If the time or date have not been changed, then there may be a problem
with the clock. Reset the clock from the SETUP menu to the current date and
time. If the problem persists, contact Ionics Instruments.
Warning 36:
The printer is out of paper and results cannot be printed. Install
paper in the printer.
Error 37:
This indicates a hardware or software error. Please note what
actions were being performed when this error occurred, and contact Ionics
Instruments.
Error 38:
The full and empty positions of the syringe pumps have been
recalibrated. This error is reported if the analyzer calibrates the syringes at
times other than power-up. Unless this error occurs frequently, no action is
required. If this error is frequent, Contact Ionics Instruments for instructions.
Error 39:
The full and empty positions of the syringe pumps cannot be
calibrated. Check to make sure that the ribbon cable located to the left of the
syringe pump PC board is plugged into the sockets. Otherwise, contact Ionics
Instruments.
Warnings 40-41:
This indicates the RAM Card has been damaged or the
format of the RAM card has been corrupted. Replace the card or follow the
procedures in the Ram Card Chapter to reformat the card.
Warning 42:
This indicates that the concentration of IC in the sample is large
compared to the TOC, and the accuracy of the measurement will be reduced.
Either use the on-line IC removal module for this sample for follow the
Содержание TOC 800
Страница 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Страница 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Страница 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Страница 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Страница 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...