Ionics Instruments
2004
Page 4-14
DLM 30007-08 Rev. A
12. Replace the reagent enclosure top cover and secure with the thumb
screws.
13. The install date will be input to the analyzer’s firmware. Record
reagent install date for input once analyzer is operational.
NOTE
Instrument firmware with the DataGuard feature differ from the basic firmware
in several ways. Access to most of the menus requires a password; some of the
menus are different, and there are some additional menus. See the DataGuard
Operation chapter for more information regarding the DataGuard feature.
Selecting the Sample Mode
The TOC analyzer can be operated in two sample modes: ON-LINE and
GRAB. The on-line sampling mode is used for continuous monitoring from a
pressurized sample stream using the on-line sample inlet system. The grab
sampling mode is used for running standards or samples from bottles or
containers. If the Multistream is installed, MULTISTREAM is a third mode; see
the Multistream Manual for details. If the analyzer is being used with the
autosampler system, the selected analyzer sample mode will be on-line.
When using an autosampler, the DataPro Autosampler Software will place the
analyzer in on-line mode and all instrument control and operation will be
achieved through the software.
To turn the analyzer on, and select the sample mode, proceed as follows:
1. Connect the AC power cord to the left-hand side of analyzer (see
Figures 4-4 and 4-5), and plug in the cord.
2. Turn the analyzer on using the main power switch. The analyzer
should power up by displaying the Analyzer Identification Screen
followed by the Main menu.
Содержание TOC 800
Страница 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Страница 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Страница 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Страница 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Страница 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...