Page 10-12
DLM 30007-08 Rev. A
Service Technician valuable information. These tests determine whether there
is a valid electronic connection.
Syringe Pump
The analyzer has the ability to perform mechanical operations associated with
the syringe pump replacement and repair. These repairs, replacements, and
menu operations should only be run with the direction of an Ionics Instruments
Service Technician.
Upon completion of a repair, scroll to MAINTENANCE on the MAIN menu and
press ENTER. Scroll to MECHANICAL and press ENTER. The following
screen will display the tests:
ENTER
CLEAR
SET DEAD VOLUME
TEST SERVO
TEST SYRINGE
LIFETIME STATS
FIGURE 10-7: MECHANICAL Menu
The SET DEAD VOLUME option is only used after the replacement of the
syringe pump or any of its components. TEST SERVO determines the
functionality of the Servo Motor. TEST SYRINGE tests the syringe for
reliability. LIFETIME STATS provides information for the Ionics Instruments
Service Technician.
Returning the Analyzer to Ionics Instruments
In some instances, it will be necessary to return the analyzer to the factory for
repairs.
Several precautions must be followed to ensure that the analyzer
is not damaged during shipment.
If the original shipping container has not been retained or heat packs are
unavailable, contact Ionics Instruments at 1-800-255-6964 or 303-444-2009 to
Содержание TOC 800
Страница 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Страница 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Страница 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Страница 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Страница 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...