Ionics Instruments
2004
Page 8-13
DLM 30007-08 Rev. A
3.
Locate the sample pump and tubing support bracket located at the back,
right-hand corner of the analyzer (see Figure 8-4). The support bracket is a
plastic piece just to the left of the pump.
4.
Remove wing nut in center of bracket and remove the piece.
Figure 8-4: Pump Tubing and Components
5. Use a 3/32" Allen wrench to loosen both of the screws on the semicircular
occlusion pressure plate on the right-hand side of the sample pump.
6. Remove the screw from the pressure plate closest to the front of the
analyzer (see Figure 8-4) and swing the plate open to expose the tubing.
7. Locate the membrane modules (circular, stainless steel pieces with tubing
extending out the top) located in the middle of the analyzer. The tubing to
be replaced is connected to the membrane module, and runs behind the
syringe pumps to the sample pump.
8. Grasp the membrane module and pull on the outlet tubing to disconnect.
While holding the tubing up, remove the tubing from around the sample
Sample Pump
Screws on
the
semicircular
occlusion
press plate
Membrane Modules
Pump Tubin
g