SITRANS LR250 with FOUNDATION FIELDBUS
36
Operating Instructions, 09/2021, A5E32221411-AH
Connecting
5
5.1
Basic safety information
NOTICE
Condensation in the device
Damage to device through formation of condensation if the temperature difference
between transportation or storage and the mounting location exceeds 20 °C (36 °F).
•
Before taking the device into operation, let the device adapt for several hours in the new
environment.
WARNING
Missing PE/ground connection
Risk of electric shock.
Depending on the device version, connect the power supply as follows:
•
Power plug
: Ensure that the used socket has a PE/ground conductor connection. Check
that the PE/ground conductor connection of the socket and power plug match each
other.
•
Connecting terminals
: Connect the terminals according to the terminal connection
diagram. First connect the PE/ground conductor.
5.2
Connecting SITRANS LR250
WARNING
Incorrect connection to power source
Risk to personnel, system and environment can result from improper power connection.
•
The DC input terminals shall be supplied from a source providing electrical isolation
between the input and output, in order to meet the applicable safety requirements of
IEC 61010-1. For example, Class 2 or Limited Energy Source.
•
All field wiring must have insulation suitable for rated voltages.