Parameter reference
SITRANS LR250 with FOUNDATION FIELDBUS
Operating Instructions, 09/2021, A5E32221411-AH
147
Reset Type
Result
Factory Defaults
a)
Default. Resets all user parameters to the manufacturer’s default settings.
Following this type of reset, complete reprogramming is required.
Standard Defaults
Resets all parameters to standard default settings.
Informational
Resets parameters such as Block Descriptor, Strategy, Device Install Date,
Device Message.
Functional
a)
Resets parameters that control device behavior and functionality (such as
Low Calibration Point).
Warm Start
Has the same effect as recycling power to the device.
FF Object Dictionary
Resets the FF standard block profile parameters (such as block tags) to
their specified defaults. This option also clears any function block parame-
ters and device schedule
b)
set by the user.
a)
The only difference between Factory Defaults and Functional reset is that Factory Defaults resets
maintenance parameters, such as device and sensor wear, calibration and maintenance timers.
Functional reset does not reset these parameters.
b)
See
Data transmission
in manual
Foundation Fieldbus for Level Instruments
(7ML19985MP01) for
further details.
To access via AMS Device Manager see Master Reset under Operation (RESOURCE) (Page 93).
To perform a reset via the handheld programmer:
1.
Press
RIGHT arrow
to open Edit Mode then scroll down to the desired reset type and
press RIGHT
arrow to select it.
2.
Press
LEFT arrow
to exit.
After performing a master reset, the device will stop measuring, the Resource and Level
Transducer Blocks will go to
Out of Service
, and the LUI will show the
Quick Start Wizard
until the device is configured.
Remaining Device Lifetime (4.2.)
Note
•
Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly
stated.
•
Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up
Maintenance/Service schedules, based on operating hours instead of a calendar-based
schedule. See also
Remaining Sensor Lifetime (4.3.)
,
Service Schedule (4.4.)
, and
Calibration Schedule (4.5.)
.
•
Performing a reset to
Factory Defaults
will reset all the Maintenance Schedule
parameters to their factory defaults.
•
The device operates in years. To view Remaining Device Lifetime parameters in hours (via
AMS Device Manager only) see
Remaining Device Lifetime
under Maintenance &
Diagnostics (RESOURCE) (Page 95)
.
The device tracks itself based on operating hours and monitors its predicted lifetime. You can
modify the expected device lifetime, set up schedules for maintenance alerts, and
acknowledge them.