Diagnosing and troubleshooting
10.2 General fault codes
SITRANS LR250 with FOUNDATION FIELDBUS
Operating Instructions, 09/2021, A5E32221411-AH
173
Code/
Icon
Meaning
Corrective Action
S: 11
Internal temperature sensor failure.
Repair required: contact your local Siemens representative.
S: 12
Internal temperature of device has exceeded
specifications: it is operating outside its tempera-
ture range.
•
Relocate device and/or lower process temperature
enough to cool device.
•
Inspect for heat-related damage and contact your local
Siemens representative if repair is required.
•
Fault code will persist until a manual reset is performed
using AMS or the LCD interface.
S: 17
Calibration interval (4.5.1.)
has
expired and has triggered a
Maintenance
Required reminder (4.5.5.)
.
Perform calibration.
S: 18
Calibration interval (4.5.1.)
has
expired and has triggered a
Maintenance
Demanded reminder (4.5.6.)
.
Perform calibration.
S: 22
Time Base Fault
Hardware problem: contact your local Siemens representa-
tive.
S: 28
Internal device failure caused by a RAM memory
error.
Repair required: contact your local Siemens representative.
S: 29
EEPROM damaged.
Repair required: contact your local Siemens representative
S: 31
Flash error.
Repair required: contact your local Siemens representative
S: 33
Factory calibration for the internal temperature
sensor has been lost.
Repair required: contact your local Siemens representative
S: 34
Factory calibration for the device has been lost.
Repair required: contact your local Siemens representative
S: 35
Factory calibration for the device has been lost.
Repair required: contact your local Siemens representative
S: 36
Unable to start microwave module.
Repair required: contact your local Siemens representative
Содержание SITRANS L
Страница 258: ...LCD menu structure SITRANS LR250 with FOUNDATION FIELDBUS 258 Operating Instructions 09 2021 A5E32221411 AH ...
Страница 259: ...LCD menu structure SITRANS LR250 with FOUNDATION FIELDBUS Operating Instructions 09 2021 A5E32221411 AH 259 ...
Страница 260: ...LCD menu structure SITRANS LR250 with FOUNDATION FIELDBUS 260 Operating Instructions 09 2021 A5E32221411 AH ...