Page 5-6
248 Operator’s Manual
Changing Beeper Option
You can turn the system beeper on or off.
To change the beeper:
from
Ready
press
#
for
menu
,
6
for
System
Setup
and
4
for
Beeper
.
Select beeper on or off.
selected
not selected
Default settings: beeper on.
Changing Communication Options
The 248 has two data ports. You can configure both ports to your
requirements. See Appendix D,
Interfacing to External Devices
for
detailed information.
To configure the data output ports:
from
Ready
press
#
for
menu
,
6
for
System Setup
and
5
for
Communications
.
Choose from:
•
CMSI
•
LIS 1
•
LIS 3.
selected
not selected
Default setting: CMSI.
Setting Security
You can protect access to the Main Menu. This guards against
unauthorized or accidental changing of setup options.
The security password only protects the Main Menu. The 248 will allow
sample and QC measurement, and will calibrate as required.
If the security option is used, an overlay appears on the Main Menu
screen prompting you to enter the password. Menu access is prevented
until the password has been entered correctly. You can return to the
Ready
screen without entering the password.
To set security:
from
Ready
press
#
for
menu
,
6
for
System Setup
and
6
for
Security
.
Enter a password. The password can be up to eight digits; you can use
the decimal point key to insert dashes.
NOTE:
If you forget the password you can use the master password
(0066838).
Default setting: security off.