Page 3-1
3 Maintaining the System
The 248 has been designed to reduce the need for maintenance to an
absolute minimum. However, careful attention to the few regular
preventative maintenance routines which are required will be more
than repaid by reliable, troublefree performance. To help with this, a
maintenance schedule is provided (see example on Page 3-3, working
copies are inside the front cover of this manual).
The maintenance frequency is based on analyzing 20 - 30 samples/day.
Increase the maintenance frequency if your laboratory analyzes more
than 30 samples/day.
We recommend using the
Disinfect
routine (Page 3-8) before carrying
out the following maintenance routines:
• replacing the pump tubing, cleaning and lubricating the rollers,
• replacing the reference sensor cassette,
• filling/replacing the pH sensor,
• replacing the
p
CO
2
and
p
O
2
sensors,
• replacing the probe and tubing, probe housing and probe protector,
• replacing the pre-heater tube.
The
Deproteinize, Condition, Prime, Disinfect
and
Stop System
routines are in the Maintenance Menu.
Maintenance should be carried out with the instrument functions
suspended, using the
Stop System
routine.
When replacing the pump tubing kits drain the 248 using the
Prime
routine.
Daily Maintenance
Equipment: Buffer Pack, (473496) as required; Wash Pack, (473497) as
required; 10% v/v bleach; clean tissues.
1.
Check levels of reagents and replace if necessary. With typical use
the reagents will need replacing every 10 to 14 days. Replace the
reagents if 30 days usage has been exceeded. Agitate the buffer
pack daily to incorporate any solution that may have condensed on
the inside of the bottles.
2.
Check the waste bottle and empty if necessary (Page 3-4).
3.
Wipe the probe sleeve, sample area, reagent compartment and
external surfaces with clean tissues moistened with 10% v/v bleach.
Do not spray into the measurement block.
Main Menu
→
Maintenance
1 Deproteinize
5 Stop System
2 Condition
3 Prime
4 Disinfect
press 1 - 5 or
*
to Exit