Maintaining the System
Page 3-13
Changing the Pump Tubing, and Cleaning and Lubricating the Rollers
BIOHAZARD:
See Appendix A,
Protecting Yourself from Biohazards
Equipment: Pump tubing kits (Cat. 673254 and 673257 or 673358);
screwdriver, supplied in the Spares Kit; mild detergent; disinfectant
(for example, Cat. 673390).
The 248 has two sets of pump tubes. The lefthand set is the sample
pump tubing and has two tubes. The righthand set is the reagent
tubing and has three tubes. For optimum performance both pump
tubing kits should be changed together.
Change the pump tubing on or before the dates shown on the label.
Changing the Pump Tubing
1.
Use the
Disinfect
routine and drain the 248, Pages 3-8 and 3-9. Stop
the 248 system, Page 3-8.
2.
Raise the front cover.
Sample Pump
Remove the tubing:
a. Remove the waste bottle.
b. Disconnect the waste cap connector from the manifold.
c. Disconnect the sample tube from the measurement block tube,
and the waste tube from the manifold.
d. Release the tension on the tubes by pulling each tube down and
to the side until the lug is clear of the tensioner.
e. Remove the pump tubing.
Figure 3-8. Removing the Sample Pump Tubing
Reagent Pump
Remove the tubing:
a. Disconnect the rubber connector from the manifold.
b. Release the tension on the tubes by pulling each tube down and
to the side until the lug is clear of the tensioner.
c. Remove the pump tubing.
Tensioner
Waste Tube
Measurement
Block Tube
Waste Cap
Connector
Lug