Precision on Controls
pCO
2
(mmHg)
Level
n
Mean
WRSD
TotSD
%CV
1
108 73.5
0.49
2.24 3.04
2
108 45.1
0.34
0.82 1.83
3
108 23.0
0.25
0.47 2.03
pO
2
(mmHg)
Level
n
Mean
WRSD
TotSD
%CV
1
108 53.8
2.78
2.49 4.63
2
108 93.8
1.16
1.88 2.00
3
108 139.5
2.21
3.32 2.38
WRSD = within-run standard deviation
Tot SD = total standard deviation
Measurement Time
Results are displayed within 45 to 120 seconds of returning the probe
(typically less than 60 seconds).
Heater
The sensor operating temperature is 37.0
°
C
±
0.15
°
C.
The pre-heater temperature is 37
°
C
±
1
°
C.
Samples
Whole blood, properly collected, refer to Page 1-7. Free from hemolysis,
and if not analyzed immediately, stored as stated on Page 1-8. Fresh
samples can be analyzed at a temperature of up to 40
°
C.
QC material, Siemens Healthcare Diagnostics recommended.
Calibration verification material, Siemens recommended.
Sample Volume
85
μ
L (syringe) nominal, minimum 60
μ
L (capillary).
Specifications
Page E-3