Page 3-2
248 Operator’s Manual
Daily Maintenance
NOTE:
Do not use any cleaning material containing alcohol as this
could cause certain components to crack.
4.
Clean the drip tray. Check it is located properly and the connector
is fitted (Page 3-22).
5.
Check there is enough paper in the printer - if the red stripe is
showing replace the paper (Page 3-23).
Weekly Maintenance
Equipment: As for daily maintenance, plus cal gas cylinder
(Cat. 001 84 146E, or 477434000 or local equivalent – contact distributor)
as required; slope gas cylinder (Cat. 001 84 147F, or 477438000 or local
equivalent – contact distributor) as required; pH fill solution (Cat.
478533) and Reference fill solution (Cat. 478822) as required.
Carry out daily maintenance and use the
Disinfect
routine (Page 3-8), and:
1.
Check the level of fill solution in the sensors. The reference sensor
should be filled to the line and the pH sensor should be almost full
with only a small bubble at the top. Refill the sensors if necessary,
Pages 3-17 and 3-19.
NOTE:
The
p
CO
2
and
p
O
2
sensors contain fill solution but cannot be
refilled. Slight discolouration of the fill solution in these sensors is normal.
2.
Check the sensors for air bubbles in the fill solution. Remove the
sensors and tap to dislodge air bubbles, Pages 3-17 and 3-19.
3.
Deproteinize and condition the sensors. (Deproteinizing and
conditioning may be prompted more frequently than once a week -
see Section 5,
Configuring the System
).
4.
Check the reference sensor for bubbles in the fill solution and for
crystal growth. If air bubbles are present, remove the sensor and
tap it to dislodge air bubbles. If crystal growth is present, remove
the sensor, empty the fill solution and rinse with deionized water,
then refill the sensor with reference sensor fill solution,
Cat. 478822. Clean off excess fill solution using lint free tissue and
deionized water. Push a clot removal line into the vent hole to clear
any fill solution crystals. See Page 3-19,
Replacing the Reference
Sensor Cassette
for details.
5.
Check gas cylinder pressure. Replace cylinder if pressure is less
than 300 psi.
Quarterly Maintenance
Equipment: As for daily/weekly maintenance, plus pump tubing kits
(Cat. 673254 and 673257 or 673358); screwdriver, supplied with the Spares
and Accessories Kit; mild detergent; drip tray (Cat. 673255) as required.
Carry out daily and weekly maintenance and:
1.
Replace the pump tubing and the pump rotor mouldings, and clean
and lubricate the pump roller assembly. Date the pump tubing
labels a maximum of 3 months ahead, Page 3-13.
NOTE:
Under heavier workload conditions it may be necessary to
replace the pump tubing more frequently.