Operating the System
Page 2-11
Analyzing QC Samples
BIOHAZARD:
See Appendix A,
Protecting Yourself From Biohazards
.
1.
If the main message shows syringe or capillary sample press
2
to
select QC.
2.
Lift the probe lever to the first position.
WARNING
Open ampules carefully.
Use ampule breakers (Cat. 47860900L) to protect your fingers.
3.
Present the QC sample to the probe and gently push the probe
sleeve back. The 248 beeps when the probe sleeve is in the correct
position and starts sampling.
4.
Hold the sample in place. The 248 beeps when sampling is complete.
5.
Remove the sample and close the probe.
Lift probe to analyze
QC sample
11:03:48
press
#
for menu
Ready
or press
1 syringe
2 capillary
Probe open
Sampling
close probe to cancel
Sampling complete
Remove sample
and close probe
Moving sample
Please wait
press
*
to cancel
Present sample
Hold sample
in place
close probe to cancel