Heater Problems
The 248 has two heater systems, one to maintain the sensor block at
37
°
C, and one to preheat samples and reagents to 37
°
C.
Sensor Block Heater
If the sensor block temperature is outside the correct limits during
sample measurement, the temperature is printed as part of the sample
result. If the heater fails the display shows Heater Failed and the 248
does not allow calibrations or sample measurement.
Pre-heater
If the pre-heater fails the display shows Heater Failed and the 248 does
not allow calibrations or sample measurement.
Use the Heater routine, Page 4-13, to determine which heater has
failed, and contact your Siemens distributor.
Using the Troubleshooting Routines
From
Ready
press
#
for
menu
and
3
for
Troubleshooting
.
Measurement Block Routine
This routine measures and displays the sensor output in mV or pA. By
comparing the readings to the values given you can see if the sensors
require maintenance, or if they should be replaced.
NOTE:
Channels that have been turned off, Page 5-5, are still measured and
displayed in this routine.
1.
From
Troubleshooting
press
1
for
Measurement Block
.
2.
Select the buffer or gas you want to run. The 248 displays the
measurement in mV/pA in addition to the normal measurement units.
3.
Compare the mV/pA readings with these values*:
7.3 Buffer pH (mV)
6.8 Buffer pH (mV)
Nominal mV
+300.0
+330.0
Total pull in range
200.0 to 400.0
24.5 to 38.0 above
7.3 buffer mV value
Action limits
<270.0 or >330.0
<27 or >35 above
7.3 buffer mV value
Main Menu
→
Troubleshooting
1 Measurement Block…
5 Roll Printer
2 Sample Flow
3 Heater
4 Electronics
press 1 - 5 or
*
to Exit
Page 4-10
248 Operator’s Manual