Page 5-2
248 Operator’s Manual
Setting Reference Ranges
Reference ranges can be set for pH,
p
CO
2
and
p
O
2
.
If a sample measurement is outside these ranges the result is flagged
on the display and on the printout.
Individual reference values may vary according to a number of factors
such as age, posture, diet, exercise and site of blood collection. We have
taken these factors into account when establishing the default values
for the 248.
Default setting:
pH
7.350 - 7.450 (35.5 - 44.7 H
+
nmol/L)
p
CO
2
35.0 - 45.0 mmHg (4.67 - 6.00 kPa)
p
O
2
80 - 100 mmHg (10.67 - 13.33 kPa)
Each laboratory should establish its own reference ranges.
To set reference ranges:
from
Ready
press
#
for
menu
,
5
for
Operating Setup
and
2
for
Reference Ranges
.
Maximum range that can be entered is the instrument measurement
range, Page E-1. Set the ranges to the maximum instrument
measurement range if you do not want to use the reference range
facility.
Choosing Units
You can choose the units of measure for parameters.
The choices are:
•
pH units or H
+
nmol/L
•
mmHg or kPa for gases
•
g/dL, g/L or mmol/L for
c
tHb.
To choose units:
from
Ready
press
#
for
menu
,
5
for
Operating Setup
and
3
for
Units.
selected
not selected
Default setting: pH units, mmHg and g/dL.
Selecting Calibration Method and Entering Gas Values
The calibration option allows you to :
•
choose how the 248 calibrates and select the maximum time
interval between calibrations (see Page 1-10 for details).
Siemens Healthcare Diagnostics recommends that the calibration
interval is set to 30 minutes.
•
enter non-Siemens gas values (see Page 3-10 for specifications of
the gas).