
Page 3-6
248 Operator’s Manual
Changing the Reagents
5.
Tilt the top of the bottle slightly away from you and slide into
position.
Figure 3-3. Positioning the Reagents
6.
Press the cap firmly onto the neck of the bottle.
7.
If you are changing the buffer bottles, date the label 30 days ahead.
Figure 3-4. Dating the Buffer Label
8.
Lower the front cover and press
*
to restart the 248.
9.
From
Maintenance
press
3
for
Prime
.
The
Prime
routine pumps the new reagent through the system.
10. When the
Prime
routine has finished, press
*
twice to exit. The 248
calibrates on return to the
Ready
screen.
Содержание RAPIDLAB 248
Страница 1: ...248 91 001N Rev T 2008 04 Operator s Manual ...
Страница 24: ......
Страница 96: ......
Страница 104: ......
Страница 118: ......