Measurement Block Routine
Cal Gas
p
CO
2
(mV)
Slope Gas
p
CO
2
(mV)
Nominal mV
–170.0
–151.0
Total pull in range
-300.0 to +100.0
12.8 to 21.1 above
cal gas mV value
Action limits
< -270.0 or >+80.0
<13.5 or >20.2 above
cal gas mV value
Cal Gas
p
O
2
(pA)
Slope Gas
p
O
2
(pA)
Nominal pA
+764.0
+80.0
Total pull in range
171 to 1961 above
–100.0 to +250.0
slope gas pA value
Action limits
<300 or >1600 above
<–50 or >200
slope gas pA value
Action:
pH - may be caused by the reference sensor, see Pages 4-1 and 4-2. If
problem persists condition or refill the pH sensor. If the problem is still
apparent, replace the pH sensor.
p
CO
2
/
p
O
2
- deproteinize/replace sensor
* at 760 mmHg atmospheric pressure, using standard gases (Cal: 5% CO
2
,
12% O
2
and Slope: 10% CO
2
and 0% O
2
)
4.
Stability – for 7.3 Buffer and Cal Gas a typical sensor will show the
following performance:
pH
p
CO
2
p
O
2
Noise - after 15 seconds
the display will not
change by more than:
0.2 mV/10 secs 1.0 mV/10 secs 2 pA/10 secs
Drift - after 15 seconds
the display will not
change unidirectionally
by more than:
1.0 mV
5.0 mV
15 pA
during the remainder of the measurement.
Instability on the pH channel may be caused by the reference sensor.
De-bubble the reference sensor and repeat the test.
5.
Press
*
to cancel the test.
Run Test Sample
1.
You can use the run test sample option to:
a. Measure a sample with a known mV value (for example, run 7.3
or 6.8 buffer as a sample).
b. Run QC samples to promote stability after installing a new
sensor.
NOTE:
You cannot use capillary samples with the Run Test Sample
routine.
Troubleshooting the System
Page 4-11