Operating the System
Page 2-5
Analyzing Capillary Samples
BIOHAZARD:
See Appendix A,
Protecting Yourself From Biohazards
.
The minimum sample volume for a capillary sample is 60
μ
L.
1.
If the main message shows syringe sample press
1
to select capillary, if
the main message shows QC sample press
2
to select capillary.
2.
Lift the probe lever to the second position.
3.
Remove the caps from the end of the capillary and carefully fit a
capillary adaptor. Slide the adaptor onto the probe, then press
*
.
The 248 beeps when you press
*
.
4.
Hold the capillary in place. The 248 beeps when sampling is complete.
5.
Remove the capillary and adaptor and close the probe.
Lift probe to analyze
capillary sample
11:03:48
press
#
for menu
Ready
or press
1 syringe
2 QC
Probe open
Present sample
then press
*
close probe to cancel
Sampling
close probe to cancel
Hold sample
in place
Sampling complete
Remove sample
and close probe
Moving sample
Please wait
press
*
to cancel