The transmitter unit includes three green LEDs for signaling the
battery status and one red LED as fault indicator (for example,
insufficient skin contact of the ECG electrodes).
(1) 3 Green LEDs (battery status)
(2) 1 Red LED (fault)
(3) Transmitter unit
Battery status and faults are also indicated on the Dot display and the
Physiological Display dialog window.
If the accumulator is not in the charging station, the green LEDs flash
regularly and simultaneously.
3 Green LEDs flash
Fully or nearly fully (2/3) charged accumulator
2 Green LEDs flash
1/3 to 2/3 fully charged accumulator
1 Green LED flashes
Nearly discharged accumulator; remaining operating duration is 1 hour
Red LED flashes (as regular as
the green LEDs)
Transmit function deactivated, unit is positioned outside the static mag-
netic field; no electrode/application fault detected
Red LED is off
Transmit function activated; unit is positioned on the patient table in
the static magnetic field; no electrode/application fault detected
Red LED flashes rapidly
PERU: electrode fault - one or more ECG electrodes are not applied cor-
rectly or fell off
PPU: application fault - pulse sensor is not applied correctly at the finger
A second set of sensors with charging station may be useful in
hospitals or radiology facilities because one PPU and PERU can be
permanently ready for operation, while a second PPU and PERU
are being charged.
The physiological signals are shown on the Dot display.
(
Control LEDs
Data display
4
Physiological imaging
88
MR System | Operator Manual
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...