
✓
A malfunction within the supervision system is detected.
✓
A corresponding error message appears at the syngo Acquisition
Workplace.
Dialog box at the syngo Acquisition Workplace
(1) Scanner hardware malfunction
Functional problem within automatic shutdown mechanism.
Please restart software of MeasRecon. If the problem persists,
please call Siemens Service immediately.
◆
Call Siemens Service.
The supervision sensitivity is potentially affected, but the
operation of your MR scanner is still possible.
Acting in case of a supervision
malfunction
3
MR system components
82
MR System | Operator Manual
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...