Only one PERU or PPU can be located in the examination room!
Two ECG and respiratory sensors in the examination room
interfere with one another in signal transmission. It is not
possible to determine the results.
Patient monitoring
systems are optional accessories and are not part
of the PMU.
The Physiological Measurement Unit may be used only for
controlling MR measurement sequences. The unit does not
replace a patient monitoring system.
The wireless PERU simultaneously acquires three ECG channels as
well as the respiratory channel of the patient.
(1) ECG leads with clips
(2) Plug for respiration cushion
(3) Transmitter unit
(4) Control LEDs
The ECG electrodes and respiratory cushion are connected to the
PERU.
To prevent skin irritations, the PERU has to be located in the
application cushion during the examination.
ECG and respiratory sensor
(PERU)
Physiological imaging
4
MAGNETOM Skyra
fit
| syngo MR E11
85
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...