(1) Table Up/Inward button
(2) Jogwheel, Center Position button
(3) Home Position button
(4) Table Down/Outward button
To operate the patient table safely and efficiently, operating
personnel must be familiar with its most important positions.
Home Position
Table is at the height of moving into the magnet bore, tabletop is
moved completely out of the magnet
Last Scan Position
If the tabletop was moved only in the horizontal direction, the tab-
letop can be returned into the position of the last measurement
Default Position
The center of the Head/Neck 20 is in the magnet isocenter
Center Position
The body region to be measured is in the magnet isocenter
Relative Position
Distance between the slice marked with the light localizer and the
magnet isocenter
For interventional examinations, a table lock function is available via
the Dot display. (
For some measurements, the tabletop is moved automatically.
3
MR system components
60
MR System | Operator Manual
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...