
Physiological imaging
Triggering methods
MR imaging procedures are sensitive to patient movement. Images
may exhibit artifacts in the form of smears when motion times - for
instance, during respiration or heartbeat - are short compared to
measurement times. In particular, this problem occurs as a result of
the patient's heartbeat during cardiac examinations or as a result of
the patient's breathing during abdominal examinations.
Two different procedures are used to avoid motion artifacts in
images: prospective triggering and retrospective gating. Both
procedures are based on the correlation between measurement and
physiological signal (ECG signal, respiratory signal, pulse signal).
Prospective triggering
During prospective triggering (or antegrade triggering), a
measurement is triggered by using a so-called trigger signal derived
from the patient's physiological signal. This signal is usually defined
based on the time period during which organ movement is as low as
possible. The trigger delay is, for example, set to the end of the
systole for certain cardiac examinations so that the measurement is
running during the akinetic diastole. For respiratory triggering during
abdominal examinations, it is recommended to start the
measurement at the end of the respiratory period.
To determine the start time for the measurement, an acquisition
window is defined based on the signal form (e.g. R-wave in ECG,
minimum of respiratory curve). For example, the size of the
acquisition window is approx. 80 % of the RR interval for ECG
measurements. The acquisition window defines the range in which
the measurement can be triggered. The trigger time is defined by the
trigger delay.
Prospective triggering can be used for ECG, pulse, or respiratory
signal curves as well as for external trigger signal curves.
4
4.1
4.1.1
Physiological imaging
4
MAGNETOM Skyra
fit
| syngo MR E11
83
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...