Look Ahead monitoring
: Prior to each measurement, the values of
the SAR quantities to be observed are calculated (always as worst-
case values) and compared with the corresponding limit values. If one
of the calculated SAR values exceeds the corresponding limit, the
measurement cannot be started. The following message appears in
the dialog window of the syngo Acquisition Workplace: SAR Limit(s)
Exceeded!
To ensure correct calculation of the SAR values, the weight of the
patient must be entered during registration.
This dialog window includes suggested changes to the examination
parameters which allow the examination to continue. In addition, a
button in the dialog window may be used to change the operating
mode.
In the First Level Controlled Operating Mode, often only minor
modifications (if any) may be required.
5
MR system operation
114
MR System | Operator Manual
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...