(1) WARNING LED
(2) POWER LED
(3) SYSTEM ON LED
LED
LEDs light up to indicate
WARNING
Error message, e.g. helium fill level is too
low
POWER
Voltage supply of MR system is satisfactory
SYSTEM ON
The MR system is switched on
1
Check the WARNING LED for alarm messages.
An alarm is present when a yellow LED lights up and/or an alarm
sounds.
2
In case of an alarm: Check the host computer for error messages.
Press the Audio Alarm Off button to silence the alarm, and notify
Siemens Service.
3
Verify that the POWER LED is green.
4
If the POWER LED is not on: Check the power supply of the MR
system.
The POWER LED is off, even though the power supply is
functioning properly?
◆
Notify Siemens Service.
After a power failure, the battery powers the circuit of the
magnet emergency shutdown for another 14 days. During this
time, the magnet can still be quenched i.e. the magnetic field
can be shut down by pressing the Magnet Stop switch in case of
emergency.
After installing remote monitoring, various error messages can be
output centrally (e.g. to the front door or gate):
Checking the LEDs
Remote monitoring
MR system components
3
MAGNETOM Skyra
fit
| syngo MR E11
73
Print No. M6-04401.621.02.02.02
Содержание MAGNETOM Skyra fit
Страница 1: ...www siemens com healthcare MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 Answers for life ...
Страница 2: ......
Страница 3: ...MAGNETOM Skyrafit Operator Manual MR System syngo MR E11 ...
Страница 6: ...Legend 6 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 122: ...6 Maintenance 122 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 128: ...This page has been intentionally left blank 128 MR System Operator Manual Print No M6 04401 621 02 02 02 ...
Страница 129: ...This page has been intentionally left blank MAGNETOM Skyrafit syngo MR E11 129 Print No M6 04401 621 02 02 02 ...
Страница 130: ...This page has been intentionally left blank 130 MR System Operator Manual Print No M6 04401 621 02 02 02 ...