6.12
Adjusting the ICR845-2
6.12.1
Adjusting the ICR845-2
To completely adjust the ICR845-2, the electrical installation must be complete and the de-
vice must be operated, see
Chapter 5.5 Performing electrical installation, Page 37
and
Chapter 6 Startup and configuration, Page 47
.
1.
Align the ICR845-2 in such a way that the red illumination area lights the code(s) on the
objects.
2.
Rotate the ICR845-2 from the plumb line to the object surface with an angle of approx.
20 so that the emitting light does not meet the code perpendicularly and does not
send back disruptive reflections. See also
Chapter 4.3.4 Selecting the rotation angle
(skew) to avoid surface reflections, Page 27
.
3.
Bring objects with codes sequentially into the ICR845-2’s field of view (trigger the
ICR845-2 for reading) and stopp the objetcs for reading the code(s). Present objects in
different positions (angles) to the ICR845-2.
4.
Check the reading result with the CLV-Setup configuration software.
5.
Align the ICR845-2 in such a way that the good read rate is between 70 and 100 %.
6.
Tighten the screws on the ICR845-2.
The ICR845-2 is aligned with the object carrying the code.