- 43 -
Main Specifications
Product Code
080130-1201
080130-1203
Model
HV-RW for aerosol
HV-RW for dioxins
Standard Suction Flow
Rate
100 or 700 L/min (quartz fiber polyurethane foam, 2)
566, 1000 L/min (glass fiber filter)
Configurable Flow Rate
Range
100 to 1200 L/min
Accuracy of Constant
Flow Rate
Within
±
5 % of the set flow rate
Suction Pressure *1
10 to 100 hPa (100 L/min)
30 to 100 hPa (700 L/min)
30 to 140 hPa (1000 L/min)
Flow Rate Detection
Differential pressure detection method
Suction Pump
Brushless blower
Filter
8"
×
10" rectangular filter
Polyurethane Foam
Dia. 90 mm
×
50 mm
Screen
Touch panel LCD (backlit)
Power Outage
Countermeasures
Continues with the prior operating conditions after power recovery
Shelter
Specified materials: aluminum (foldable)
Operating Temperature
and Humidity Range
0 to 40
°
C 10 to 90 %RH (with no condensation)
Power Supply
AC100 V (90 V to 132V), AC 220 V (180 V to 264 V) 50/60 Hz
Current Protection
Function
15 A Circuit protector
Power Consumption *2
100 L/min
:
0.4A (quartz fiber polyurethane foam, 2)
700 L/min
:
3.7A (quartz fiber polyurethane foam, 2)
700 L/min
:
2.3A (quartz fiber filter)
1000 L/min
:
4.1A (quartz fiber filter)
1000 L/min
:
3.3A ( glass fiber filter)
Dimension
Usage conditions: 575 (W)
×
575 (D)
×
1420 (H) mm,
Sampling port height: 1200 mm (from installation surface)
Folded up: 575 (W)
×
575 (D)
×
960 (H) mm
Weight
Approx. 31 kg
Accessories
Filter case, 1
Through pipe, 1
Filter case, 1
Shuttle tube, 1
Quartz fiber filters, 10
polyurethane foams, 10
*1 Operation under 1 atmospheric pressure assumed. Performance may be inferior under low
atmospheric conditions at high altitudes.
*2 This is the initial value at time of shipment, at AC100V. Power consumption may exceed double the