- 36 -
The factory default span value is 1.000, and the zero value is 000.0.
Calibration Mode
In the menu screen, press CAL to switch to calibration mode. This section only describes
the operating methods in the calibration mode. When performing an actual calibration, see the
section on calibration methods. (See page 38.)
[SPAN] key ········ Pressing this key will proceed to the respective span value entry screens.
[ZERO] key ······· Pressing this key will proceed to the respective zero value entry screens.
Leak check ········ Pressing this key will proceed to the leak check screen. (See page 37.)
Blower operating time ······ Displays the operating time for the blower mounted in this product.
[MENU] key ······· Pressing this key will return to the menu screen.
[ESC] key ········· Pressing this key will return to the main screen.
The respective span value/zero value entry
screens will be as in the figure at right. Press a
numerical region to proceed to the numerical
value entry screen.
When numerical values have been entered, the
date of entry is recorded, and the system returns
to this screen.
Press ESC to return to the menu screen.
In the numerical value entry screen, press the
numerical keys to enter the desired numerical
values.
Press
to confirm the numerical values as the
setting values and return to the respective span
value/zero value entry screens.
Press ESC to cancel entry and return to the
respective span value/zero value entry screens.
Press CLR to set the numerical value to "0"
Press BS to delete the last figure entered.
DEL is not available here.
Blower
operating time
MENU key
ESC key
Leak check
Flow rate
Atmospheric pressure
Atmospheric
temperature
ZERO key
SPAN key