- 32 -
The blower will automatically stop if the internal temperature reaches or exceeds 95
°
C.
In this case, [RPM] will be displayed as an error indication.
Detailed Error Display
TRBL is displayed when an error has occurred. Press TRBL to view detailed error
information.
The error details can be viewed as shown below. The last 5 errors are logged in sequence
from the oldest, and can be viewed by pressing the respective number keys. The 6th and
subsequent errors are not logged. If there are no errors, "NO TRBL" is displayed. In the
example in the figure below, flow rate errors have occurred 5 times, atmospheric pressure
errors 4 times, and atmospheric temperature errors 3 times, but no power errors have
occurred.
The threshold for each error can be configured in the error setting screen. (See page 35.)
A blower error key is only displayed if a blower error has occurred.
Press the respective numerical keys to display error detail screens as in the figure.
Press BLOWER to display screen will appear as in the figure.
Numerical keys
[ESC] key
Flow rate error
Blower error key
Atmospheric
pressure error
Atmospheric
temperature error
Power error
Error details and error values
Error end time
Error start time
RPM: rotational error
ERR: Other errors
Time of error occurrence