- 25 -
When the power is turned ON, a power error is displayed in the error item field. This is
because the product inadvertently detects a power outage when the power is turned
OFF, and does not indicate a malfunction. The error will disappear when the
[START/STOP]
button is pressed, and operation starts.
If the power error is displayed after operation ends, a power outage likely occurred.
Error Displays
If an error occurs during operation, this product will log the error (See page 32.). Unless the
error involves the blower however, the blower will not stop, and operation will continue.
If an error occurs, it will be indicated onscreen.
If the error is blinking, this indicates that it is occurring right now, whereas if the error is lit, this
indicates that it occurred previously. A blower error is always lit.
The internal battery run-down icon blinks when the battery power is low, and is lit when the
battery is completely dead.
If a blower error occurs, the screen display at right
will appear, with the screen backlit in red. The
blower will stop and operations will be aborted.
This screen will persist until ESC is pressed.
Error Item
FL: Flow rate error
PR: Pressure error
TE: Temperature error
AC: Power error
BL: Blower error
Internal battery
is dead.
(See page 42.)