SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 45
Further details are given in Appendix A.7.
Figure 40 Fuel shut-off/hydrogen relay output connections
5.3.5 Ethernet connection
WARNING
The digital communications terminals are separated from the
analyser mains circuits by reinforced insulation. The terminals
must only be connected to circuits that are themselves separated
from mains voltage by at least reinforced insulation.
CAUTION
To comply with EMC requirements, the cable connected to the Ethernet
output terminal shall not be longer than 30 m, nor shall it leave the
building in which the analyser is installed without suitable isolation.
Connect the cable to the terminal shown in Figure 1.
Section 2.9 provides information on the rating and size of cable.
Содержание SERVOPRO FID
Страница 2: ...This page intentionally blank...