![Servomex SERVOPRO FID Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/servomex/servopro-fid/servopro-fid_operators-manual_1247061040.webp)
SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 38
This menu is used to generate flow tables used during manufacturing or if a
restrictor is changed. Pressure set point of each regulator can be changed using
this menu.
4.8.10
Other parameters
This menu is used to configure the Ethernet password that is used to get connected
to the analyser using Ethernet network and the lock range feature.
Figure 35 – Other parameters menu
4.8.10.1
Lock range
This feature enables the user to lock the operating range. To lock a range, the
analyser must be in manual ranging mode. This feature is used when the customer
does not monitor the actual operating range. Because the weight of the 4-20 mA
output is affected by the operating range (see Appendix A5), locking the range
makes sure that the reading on the 4-20 mA is not false.
<-- <<<HIDDEN 15: Other parameters >>>
1) Ethernet Password: KONTROL
2) Lock Range: YES
Содержание SERVOPRO FID
Страница 2: ...This page intentionally blank...