SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 34
4.8.3 Clock
The clock menu allows configuring the date and time of analyser. This is used to
report error messages.
Figure 23 – Time and date menu
To configure the time, press 1 and then enter hour, minute and second. Between
each value, press E to validate. Do the same for the date.
4.8.4 Log Amp
This menu allows configuring the amplifier board.
Figure 24 - Log amplifier menu
Figure 25
Parameter
Range
Description
Pre-amp
1-1000
First stage gain of the conditioning module
Gain
1-4095
Gain of the conditioning module
Offset
0-10
Offset in volt
.
Counts
0-16777216
Amplifier signal.
9
From 0 to 4.9999 the offset is negative and from 5.0001 and 10 the offset is
positive.
<-- <<<HIDDEN 3: Time and Date>>>
1) Time HH:MM:SS 12:59:12
2) Date DD/MM/YYYY 13/01/2009
<-- <<<HIDDEN 4: Log amplifier>>>
1) Preamp: 10 Counts: 3298034
2) Gain: 10.0000
3) Offset: 5.0300