SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 39
5 INSTALLATION AND SETUP
5.1 Installation and switch-on
WARNING
Ensure that cables and tubes that you connect to the SERVOPRO
FID are routed so that they do not present a trip hazard.
WARNING
Fuel and sample gas may be explosive. Ensure that the external
connections are leak free at full operating pressure.
WARNING
Analyser uses Hydrogen fuel gas and may be explosive. Ensure
that the analyser is installed in a forced ventilated cabinet and
installation complies with all applicable site safety requirements.
CAUTION
Do not use the SERVOPRO FID in an area subject to high levels of
vibration or sudden jolts. If you do, sample measurement may not be
accurate, or the SERVOPRO FID may be damaged.
1.
Place the SERVOPRO FID in a suitable operating location, within easy
reach of a suitable electrical supply outlet. Fit the power cord to the
SERVOPRO FID’s power inlet.
2.
Fit the power cord to a suitable electrical outlet.
3.
Connect air, fuel and sample inlet. For all gas line connections (including
calibration gas and sample gas), we recommend the use of 1/8’’ stainless
steel tubing in full length, with no fittings. You must avoid pipe thread
connections: they are usually sealed with Teflon tape and some particles
can get into lines. By using compression-type tube fittings, the venturi
aspiration of outside contaminants is virtually eliminated. Tube quality is
often overlooked. For a 1/8’’ O.D. tube, use a minimum wall thickness of
0.028’’. The tube must be purchased to meet ASTM69 specifications.
Inferior quality tubes may have irregularities on their circumference,
WARNING
Carbon dioxide is an asphyxiant. Ensure that the
Installation location is vented to ensure a CO2 rate of 1.1 l/min
can be diluted to levels below 5000 ppm