SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 22
4.4 Run menu
During normal operation, the analyser should be in this menu. It allows the user to
retrieve the important data like actual impurity, status, alarms, flows and
temperatures.
Figure 3 – Run menu
The SERVOPRO FID comes with three ranges. The active range can be changed
using the keypad. To change range, use key 1, 2 or 3 on the keypad. Key 1 will
select range 1 and so on.
Range can only be changed when ranging mode is set to manual.
This menu also has a flame indicator. When the flame is off, “Flame off” is
displayed instead of measurement. When flame is off, “Press F1 to ignite” is
displayed as show in figure 3.
It is normal that “PRESS F1 to ignite” is not displayed when errors
are active.
The reading can be replaced by three types of error. If the analyser is in over
range, the field will be replaced by over scale and if the analyser is underscale, the
field will be replaced by underscale. See section 9.2 for more information. The
other possibility is flame off as already explained.
Alarms are also reported in this menu. They are displayed below the fuel field.
When there is no alarm, OK is displayed. When alarm one is active, Alarm #1 is
displayed and when both alarms are on, Alarm #1,#2 is displayed. See section 6.2
for more information regarding how the alarms are working.
Alarms information is only displayed when this option is
selected.
Measure: Flame off
Smp: 30 CC Air: 300 CC Fuel: 100 CC
Range:0-1 Status: Error
FID temp: 100C PRESS F1 to ignite