![Servomex SERVOPRO FID Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/servomex/servopro-fid/servopro-fid_operators-manual_1247061018.webp)
SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 16
2.9 milliAmp (mA) output
Maximum load
500 Ω
Connector type
Terminal block
Isolation voltage
500 V RMS
Output range
Normal sample measurement
Under range
Over range
4 to 20 mA
Less than 4 mA
Over than 20 mA (24 mA maximum)
Cable requirements
Type
Maximum size
Multi-strand twisted pair with overall
screen
0.82 mm
2
(18 AWG)
Table 10 : MilliAmp (mA) output specifications
2.10 Digital outputs
Total maximum current
1 Amp
Connector type
Terminal block
Maximum voltage
24 VDC
Cable requirements
Type
Maximum size
Multi-core cable with overall screen
0.82 mm
2
(18 AWG)
Table 11 : Digital output specifications
Содержание SERVOPRO FID
Страница 2: ...This page intentionally blank...