![Servomex SERVOPRO FID Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/servomex/servopro-fid/servopro-fid_operators-manual_1247061027.webp)
SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 25
occurs, 4-20 mA analogue
output is set below 4 mA.
3.
HIGH
: If a system error
occurs, 4-20 mA analogue
output is set above 20 mA.
Alarm #1
ppm
0-1000
Alarm #1 set point. If the
calculated impurity is over this set
point, an alarm #1 is generated.
Alarm #1 must always be smaller
then alarm #2. See section 6.2 for
more information.
Alarm #2
ppm
0-1000
Alarm #2 set point. If the
calculated impurity is over this set
point, an alarm #2 is generated.
See section 6.2 for more
information.
Auto-ignition
1-5
This is the number of times that
the analyser will try to restart the
flame if the flame goes out. See
section 6.4.1.
Table 16 : Configuration parameters
Parameter
Range
Description
Ethernet
enabled
YES
: Analyser can be remotely controlled
NO
: It is impossible to get connected to the
SERVOPRO FID.
DHCP
enabled
YES
: DHCP is on.
NO
: SERVOPRO FID uses a static IP address.
IP
0-255
If DHCP is enabled, the field is set read-only.
Otherwise, user can set the static IP.
Mask
0-255
If DHCP is enabled, the field is set read-only.
Otherwise, user can set the network mask.
Gateway
0-255
If DHCP is enabled, the field is set read-only.
Otherwise, user can set the network gateway.
MAC
This field is read-only. This is the analyser’s MAC
address.
Table 17 : Connectivity parameters
Содержание SERVOPRO FID