70 DF-760E
User Interface
Figure 55: Purge Cancel
7.5.9.4 Pump Capacity Test
The pump capacity test can be used to determine the ultimate vacuum that the aspirator or
pump is capable of pulling. Proper flow through the analyzer is a direct result of the proper
balance between the pressure on the analyzer inlet and the vacuum at the analyzer outlet. If
either is incorrect, the flow will suffer and the moisture readings will not accurately reflect
the condition of the process gas.
Figure 56: Pump Capacity Test
The test is automatic in nature, in that once started, the analyzer isolates the moisture cell
by closing the upstream and downstream valves and then the vacuum is monitored for 20
seconds and is displayed on the screen. Any pressure lower than 115 Torr is considered
acceptable. If the aspirator or pump is unable to pull 115 Torr, a failure is indicated by the
appearance of Figure 57 and the recommendation to rebuild the pump or check for leaks.
At the end of the 20 second test, whether the aspirator or pump passes or fails, the user
must hit Escape to return to the main display.
After the pump test is complete, an entry is automatically put in the pump capacity test log
as shown in Figure 58. See page 61 for additional information on the pump test log.
Figure 57: Pump Pressure Failure
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...