Sample Gas Preparation and Delivery
8.4.3 Pressure Effects on Oxygen Sensor Performance
If the analyzer is not vented to atmosphere, the oxygen sensor pressure is influenced by the
conditions downstream of the analyzer. A recalibration under your operating conditions
may be desirable to remain within the stated accuracy specifications. However, in most
cases the error introduced is relatively small, and may not affect the process application.
Sample gas line lengths, fittings and bends should be kept to a minimum to maintain low
pressure drops. Larger diameter tubing and fittings reduce pressure drop and also lengthen
response time. In general, 1/8-inch tubing should be limited to 15-foot runs; longer runs
should be made with 1/4-inch tubing.
8.4.4 Sample Outlet Backpressure Effects
It is always recommended to vent the analyzer to atmospheric pressure. However, if a
sample vent or return line is used, attention must be given to maintain a low and consistent
backpressure so as not to affect the flow rate.
The allowable backpressure on the
oxygen
sensor is ±1 psig. If variations in the vent line
pressure are expected, a sub-atmospheric backpressure regulator should be installed on the
vent line to maintain an even backpressure on the analyzer. Consider the regulator’s
pressure drop (typically 1 psi) when designing the sample vent system in order to stay
within the ±1 psig pressure limits at the sensor.
When not venting the analyzer to atmosphere, it is also suggested to install a fairly high
resolution pressure gauge immediately at the analyzer outlet.
8.4.5 Flammable Sample Gas
There is nothing within the analyzer sample system that can ignite a flammable sample
gas. However, it is critical to ensure that the sample gas does not escape from the sample
system into the analyzer enclosure, or the room, where ignition is possible.
Also, the analyzer enclosure can be purged with nitrogen, or the entire Analyzer can be
mounted in a purged enclosure, so that any sample gas that escapes the plumbing is
diluted. Servomex does not provide an analyzer equipped with these protective features.
Note that these methods may not be compliant with local or national requirements in your
country. Ensure that you comply with all necessary regulations and standards.
8.4.6 Sample Gas Temperature
Gas temperature should not exceed 50 °C (122° F), nor should it fall below 0° C (32° F).
Gas temperature can be controlled by passing the gas through 5 to 10 feet of metal tubing
that is within the recommended sample temperature. Because of its low thermal mass, the
gas sample quickly reaches the gas sample line temperature.
The analyzer has software to correct the sensor output for sensor temperature changes.
Temperature compensation adjustments apply to temperature drift only when the oxygen
level is below 10 ppb. Ideally, the analyzer should be operated at a nominal temperature of
70° F. Calibration temperature should be close to operating temperature. If the analyzer is
to be operated at an average ambient temperature outside 65° F to 80° F, it should be
recalibrated at the operating temperature for optimal performance.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...