Service DF-745
135
9.5
Troubleshooting Guide for the NanoTrace Dual
Analyzer
9.5.1.1 Oxygen Troubleshooting
Observation
Possible Remedy
(see Key below)
Analyzer reads Over-Range
Y, C, Q, I
Analyzer spikes excessively when moved using portable feature
B, J, I
Analyzer output has unacceptable peak-to-peak noise
J, H, X, I
Zero baseline gradually drifting positive
G, A, B, C, D, Q, H, AB, I
Zero baseline gradually drifting negative
P
Zero baseline high, but stable (> 15 ppb above factory zero)
G, A, B, C, D, E, Q, AB, I
Very slow analyzer purge down (doesn’t drop below 10 ppb in 7 days)
G, A, B, C, D, E, Q, H, AB, I
Zero baseline drifting up and down (exclusive of temperature)
H, Q, I
Repetitive negative spiking
J, X, Z, A, B, C, D, E, Q, H, I, AA
Repetitive positive spiking
J, X, Z
O
2
reading is drifting excessively with ambient temperature (> 0.3 ppb/C)
E, G, P, R, AB, I
Electrolyte Condition Alarm ON
A, N, C, D, H, I
O
2
reading does not decrease upon switch to on-board Delta F purifier
(Assumes sample gas contains some O
2
)
O
Span reading is unacceptably high (> 50% high)
T, V, S, I
Span reading is unacceptably low (> 50% low)
T, S, I, H
Unacceptably Slow Speed of Response
L, G, H, I
Analyzer indicates high temperature
AB, I
9.5.1.2 Key:
A
Add Delta F Replenishment Solution if level is near or below “MIN” mark
B
Remove some electrolyte if level is near or above “MAX” mark
C
Measure applied voltages on electrode pairs:
Sensor Electrodes (wht/yel* and wht/red/blk) 1.300 ± 0.005 VDC
Secondary Electrodes (wht/blue* and wht/red) 2.1 ± 0.3 VDC
Stablex Electrodes (white* and blue) 1.55 ± 0.005 VDC
* is the common lead of the voltmeter.
D
Measure the DC currents on electrode pairs:
Secondary Electrodes 2.0 ± 0.2 mADC (disconnect wht/red wire at reservoir terminal, and insert ammeter between
wht/red wire and reservoir terminal)
Stablex Electrodes < 13 uADC
(disconnect white wire at reservoir terminal, and insert ammeter between white wire and reservoir terminal)
E
Check sensor temperature in Diagnostics Menu. It should be approx. equal to the current ambient temperature.
G
Perform the low flow leak test: Obtain stable oxygen readings at flow = 2.0 scfh and flow = 1.0 scfh. The reading at flow
= 1.0 scfh should be no more than 2 ppb higher than that at 2.0 scfh. Locate and fix any ambient leaks upstream of the
analyzer. See page 120.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...