User Interface DF-760E
93
7.6.6.5 View Oxygen Over Temperature Log
Figure 93: Oxygen Over Temperature Log
7.6.7 Oxygen Alarm SetUp
Figure 94: Oxygen Alarm SetUp Menu
The oxygen analyzer includes a total of eight alarms. The four oxygen concentration
alarms can be user controlled to activate up to four optional relays. High and low setpoints
as well as deadbands are user-set.
The Temperature, Low Flow and Electrolyte Condition alarms all relate to the operation of
the oxygen sensor.
In addition, an alarm can be programmed to indicate whether the oxygen sensor has been
turned off, either manually or automatically.
NOTE: When any hydrogen is included in the background gas matrix (see discussion
regarding GSF on page 103 for additional information), an additional alarm is enabled to
monitor the cabinet exhaust fan status. If the system detects a failure in the operation of the
cabinet exhaust fans while operating in a hydrogen background, the system immediately
isolates the oxygen and moisture cells until the fan is repaired. A warning describing this
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...